현장검사 분석 기기인 ABL90 FLEX PLUS의 크레아티닌 분석능 평가
- 제목
- Performance Evaluation of the ABL90 FLEX PLUS Point-of-Care Analyzer in Measuring Creatinine Levels
- 저자
- 이승재
; 안효준
; 유종하
- 키워드 (영문)
- creatinine; abl90 flex plus; point-of-care testing; kidney function tests
- 발행연도
- 2020-07
- 발행기관
- 대한임상화학회
- 유형
- Article
- 초록
- Background: To diagnose acute renal failure, creatinine levels in whole blood are typically assessed using point-of-care testing (POCT) methods. The present study aimed to evaluate the performance of a newly developed POCT blood gas analyzer, the ABL90 FLEX PLUS (Radiometer, Denmark), which can measure creatinine in blood.
Methods: Precision and linearity of the ABL90 FLEX PLUS were evaluated and compared with those of the Beckman Coulter AU5800 (Beckman Coulter, USA), according to the CLSI guidelines for creatinine measurement performance.
Results: For the ABL90 FLEX PLUS, the total imprecision (%CV) levels of two control materials were measured to be 0.0% and 0.8%, while linearity was evaluated, with the R2 value measured to be 0.9993 (0.4-8.4 mg/dL). Compared to the AU5800, the ABL90 FLEX PLUS correlation coefficient (r) was found to be 0.989. The 95% limits of agreement were determined to be -0.649 and 0.643 mg/dL (-18.8% and 17.8%).
Conclusions: The ABL90 FLEX PLUS exhibited good performance for creatinine test. This indicates that the ABL90 FLEX PLUS can be potentially useful in clinical laboratories.
- 저널명
- Laboratory Medicine Online
- 저널정보
- (2020-07). Laboratory Medicine Online, Vol.10(3), 202–206
- ISSN
- 2093-6338
- DOI
- 10.3343/lmo.2020.10.3.202
- 공개 및 라이선스
-
- 파일 목록
-
Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.