Evaluation of the Reliability of Automated Colistin Susceptibility Assays of Pseudomonas aeruginosa
- 저자
- Young Ah Kim
; Junglim Kim
; Young Jun Son
; Sang Sun Lee
- 키워드 (영문)
- colistin; automated susceptibility assay; broth microdilution; pseudomonas aeruginosa
- 발행연도
- 2022-01
- 발행기관
- CrossRef
- 유형
- Article
- 초록
- Background: With the increase in multidrug-resistant (MDR) gram-negative rods, colistin treatment is increasingly required. Reliable susceptibility testing is important for managing infectious patients with MDR Pseudomonas aeruginosa . This study evaluated the reliability of an automated colistin susceptibility assay.
Methods: In total, 195 non-duplicated P. aeruginosa blood isolates were collected from nine sentinel hospitals in Korea in 2020. Colistin susceptibility with Sensititre KNIHCOL Trek diagnostic system (Thermo Fisher Scientific, England), a commercial broth microdilution (BMD), was compared with that of the gram-negative MIC panel type 44 card (Beckman Coulter, USA). Isolates with discrepant results were reconfirmed using AST-N225 (bioMérieux SA, France).
Results: Categorical agreement was 92.8%, and nine colistin-resistant isolates with BMD were colistin-intermediate with two automated colistin susceptibility assays.
Conclusions: The failure to detect colistin-resistant P. aeruginosa could not be safely excluded with an automated colistin susceptibility assay.
- 저널명
- Laboratory Medicine and Quality Assurance
- 저널정보
- (2022-01). Laboratory Medicine and Quality Assurance, Vol.44(1), 36–39
- ISSN
- 2384-2458
- DOI
- 10.15263/jlmqa.2022.44.1.36
- 공개 및 라이선스
-
- 파일 목록
-